예랑솔루션

PRODUCT

고객님께 적합하고 효율적인 최선의 솔루션을 제안하겠습니다.

gwinstek>>Gwinstek>>LCR-6000 시리즈

  • LCR_6000.png
LCR-6000 시리즈 고정밀 LCR 미터
모델명LCR-6300, LCR-6200 외 테스트주파수10Hz~2kHz/20kHz/100kHz/200kHz/300kHz 기본정확도0.05%
제품상세정보

주요 특징

10Hz~2kHz/20kHz/100kHz/200kHz/300kHz
0.05% 기본 정확도
연속적인 테스트 주파수 설정
3.5인치 컬러 LCD
측정 속도 : 최대 25ms
Full/Spot 주파수 OPEN/SHORT 캘리브레이션
16개 측정 조합 + 2개 파라미터 모니터링

DCR 측정 및 내부 DC 바이어스(±2.5V)
PASS/FAIL 판정 기능
ALC(Auto Level Control) 기능
BIN 기능 : 9BIN + 1AUX
10개 스텝 테스트 기능(주파수/전압/전류)
RS-232C, Handler, USB(Host) 인터페이스
2U, 1/2 랙 사이즈 설계

Consecutive frequency and Convenient zero function

The LCR-6000 series, within the provided frequency range, features consecutive and adjustable frequency capability which allows users to conduct measurement and analysis on components with the most genuine frequency requirements. For OPEN/SHORT fixture compensation function, the LCR-6000 series is equipped with full frequency range zero and spot zero selections. After executing full frequency range zero, users, under the conditions of not turning off the power and not changing test fixture, can freely change test frequency for the LCR-6000 series to execute component measurements that tremendously saves time in repeatedly zeroing test fixture after changing frequency.

01
 Consecutive and adjustable frequency
 Freely input frequency within provided frequency range

02
Selectable fixture zeroing methods
 Full frequency range zero or spot zero

Rich and diverse information display

The measurement result display of the LCR-6000 series not only reveals primary and secondary measurement parameters but also includes two monitoring parameters. Therefore, four DUT related parameters can be simultaneously shown on the display screen to save time if repeated measurements are required. With respect to display screen, the LCR-6000 series features diverse display to meet users’ observation requirements. For instance, MEAS display shows setting parameters and measurement results at the same time; ENLARGE display focuses on measurement results and PASS/FAIL judgment is available, which is conducive to assist engineers to swiftly obtain the validity of measurement results.

01
 MEAS display
Parameter setting and four measurement parameters 

02
ENLARGE display
Enlarge measurement results and include PASS/FAIL judgment

10 points listed tests and PC software

The LCR-6000 series provides the 10 points listed test function, which allows users to define a set of DUT measurement parameters (such as Cs- Rs) and to set 10 test criteria of category (either by frequency or by voltage or by current) but different values to conduct measurements. Through this function, users can rapidly and clearly obtain DUT’s characteristic variation trend to determine the adaptability of DUT’s practical applications. The measurement results can be recorded directly in the internal memory and be transferred to the PC through USB. The LCR-6000 series also provides free PC software (maximum 1,000 points listed tests) in order to satisfy users’ analytical requirements on delicate variation.

01
 Listed tests
Variation criteria based upon frequency or voltage/current

02
On software ─ characteristic curve
Provide more delicate characteristic variation trend

Diverse ancillary measurement functions

To satisfy the diverse measurement application requirements for different components and materials, the LCR-6000 series collocates with many auxiliary measurement functions. For capacitor measurement, Automatic Level Control (ALC) is mainly for component which requires a constant or rated test voltage such as multi-layer ceramic capacitor (MLCC). An internal D.C. bias voltage (±2.5V, internal) is allowing simulating A.C. and D.C. coexistence to learn capacitance variation. For inductor measurement, the D.C. resistance measurement function is to validate D.C. resistance characteristics. Additional, the LCZ function is to quickly identify components ’characteristics. When the function is activated, the LCR-6000 series will automatically determine DUTs’ characteristics and reveal the optimum parameters to show the measurement results.

01
Automatic level control
Ideal for measuring components with voltage requirements

02
Internal bias (±2.5V adjustable)
Ideal for capacitive components’ characteristic tests  

01
 D.C. resistance measurement
Ideal for inductive components’ D.C. characteristics verification

 

Standard Interface

For interface connectivity, the LCR-6000 series comes equipped with handler interface and RS-232C interface. Handler outputs 10 BIN (9BIN, AUX: 1BIN) sorting results that is best for external connection control, for instance, connecting to a sorting machine to conduct components’ sorting operation. RS-232C is suitable for remote control and measurement results retrieval. The PC gives commands to control settings or to read measurement results so as to achieve the requirements of verifying automotive applications.

02

TEST FREQUENCY

 

LCR-6300: 10Hz~300kHz(±0.01%) (4 digits resolution)
LCR-6200: 10Hz~200kHz(±0.01%) (4 digits resolution)
LCR-6100: 10Hz~100kHz(±0.01%) (4 digits resolution)
LCR-6020: 10Hz~ 20kHz(±0.01%) (4 digits resolution)
LCR-6002: 10Hz~  2kHz(±0.01%) (4 digits resolution)

OUTPUT IMPEDANCE

 

30Ω / 50Ω / 100Ω selectable

BASIC ACCURACY

Slow / Med

0.05%

Fast

0.1%

TEST SPEED

 

FAST:25ms / MED:100ms / SLOW:333ms

TEST SIGNAL LEVELS

Voltage

10.00mV- 2.00V (±10%)   CV :10.00mV- 2.00V(±6%)

Current

100.0uA- 20.00mA (±10%) CC :100.0uA- 20.00mA(±6%) (@2VMax)

DC BIAS

Internal

±2.5V (0.5%+0.005V)

DISPLAYRANGE

R, X, |Z|

G, B, |Y|

L

C

D

Q

θd

θr

DCR

Δ%

0.00001Ω ~ 99.9999MΩ

0.01nS ~ 999.999S

0.00001uH ~ 9999.99H

0.00001pF ~ 9999.99mF

0.00001 ~ 9.99999

0.00001 ~ 99999.9

-179.999° ~ 179.999°

-3.14159 ~ 3.14159

0.00001Ω ~ 99.9999MΩ

-99999% ~ 99999%

TEST MODE

16 Combinations

Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-Rp, Lp-Q, Ls-Rs, Ls-Q,
Rs-Q, Rp-Q, R-X, Z-θr, Z-θd, Z-D, Z-Q, DCR

Monitor parameter (2 selectable)

Z, D, Q, Vac, Iac, Δ, Δ%, θr, θd, R, X, G, B, Y

LISTED MODE

 

10 steps

BIN FUNCTION

 

Comparator (9BIN,AUX:1BIN)

MEMORY

INT – panel setting

10 file name

INT – measured data

10000 Data(.csv)

USB storage

10 file name for setting, 9999 file name for data, 999 Log file for LCD screen

OTHER FUNCTION

Auto Level Control (ALC)

ON/OFF

Average

1~256 times

Trigger

INT / MAN / EXT / BUS

Delay

0ms~60s

Judgment

PASS / FAIL

Auto LCZ

ON / OFF

Screen Capture

Saving in to USB (Bmp form)

DISPLAY

 

3.5'LCD, RGB color (320x240)

INTERFACE

 

RS-232(SCPI), Handler, USB Host

POWER SOURCE

 

AC 90V-250V, 50-60Hz, Max. 30VA

DIMENSION & WEIGHT

 

265(W)x107(H)x312(D) mm, Approx.3kg

 

TOP