PRODUCT
고객님께 적합하고 효율적인 최선의 솔루션을 제안하겠습니다.
tektronix>>고전압차동프로브>>THDP0100
Tektronix offers a broad portfolio of high-voltage probing solutions that enable users to safely and accurately make floating measurements.
Key features
-
Overrange Indicator
-
Safety Certified
-
Switchable Attenuation
-
Switchable bandwidth limit
Applications
-
Floating measurements
-
Switching power supply design
-
Motor drive design
-
Electronic ballast design
-
CRT display desingn
-
Power converter desigh and service
-
Power device evaluation
Safe high-voltage probe solutions
The THDP0100 has the largest differential dynamic range capability from Tektronix, allowing users to safely measure up to ±6000 V. The THDP0100 supports bandwidths up to 100 MHz and slew rates up to 2500 V/ns at 1/1000 gain. This probe is supplied with two sizes of hook tips and has an overrange visual and audible indicator which warns the user when they are exceeding the linaer range of the probe.
Specification
Attenuation |
100X / 1000X |
Differential Voltage |
1000X : ±6000 V 100X : ±600 V |
Common Mode Voltage |
±6000 V |
Maximum Input Voltage-to-Earth |
2300 V CATⅠ 1000 V CATⅢ |
Bandwidth |
100 MHz |
Rise Time |
<3.5 ns |
Slew Rate |
<2500 V/ns at 1/1000 gain |
Input Impedance at the Probe Tip |
40 MΩ || <2.5 pF |
Typical CMRR |
DC : >-80 dB 100 kHz : >-60 dB 3.2 MHz : >-30 dB 100 MHz : >-26 dB |
Cable Length |
1.5 m |