예랑솔루션

PRODUCT

고객님께 적합하고 효율적인 최선의 솔루션을 제안하겠습니다.

tektronix>>고전압차동프로브>>THDP0200

  • 2034374844_JNk8I0tZ_THDP0200.jpg
THDP0200 Safe high-voltage probe solutions
모델명THDP0200 감쇠50X/500X 대역폭200MHz 임피던스프로브 팁: 10MΩ, <2pF 최대전압1000V CAT II, 600V CAT III 스코프DPO/MSO/MDP시리즈
제품상세정보

Tektronix offers a broad portfolio of high-voltage probing solutions that enable users to safely and accurately make floating measurements.

 


 

Key features

  • Overrange Indicator
  • Safety Certified
  • Switchable Attenuation
  • Switchable bandwidth limit

 

Applications

  • Floating measurements
  • Switching power supply design
  • Motor drive design
  • Electronic ballast design
  • CRT display desingn
  • Power converter desigh and service
  • Power device evaluation

 


 

Safe high-voltage probe solutions

 

The THDP0200 is a active differential probe that are capable of safely measuring differential voltages up to ±1500 V. The probe is effective in making measurements in IGBT circuits such as motor drives and power converters. The THDP0200 supports bandwidths up to 200 MHz and slew rates up to 275 V/ns at 1/250 gain.

 


 

Connectivity options

The TMDP and THDP Series probes are equipped with the TekVPI® interface which allows smart communication between the oscilloscope and probe. Pushing the probe menu button will launch the probe control menu on the oscilloscope display providing access to all relevant probe settings and controls. These probes are designed to operate on TekVPI® oscilloscopes without requiring the TPA-BNC adapter.

 


 

Specifications

 

 Attenuation

 50X / 500X

 Differential Voltage

 500X : ±1500 V

 50X   : ±150 V

 Common Mode Voltage

 ±1500 V

 Maximum Input Voltage-to-Earth

 1000 V CATⅡ

 600 V CATⅢ

 Bandwidth

 200 MHz

 Rise Time

 <1.8 ns

 Slew Rate

 <650 V/ns at 1/500 gain

 Input Impedance at the Probe Tip

 10 MΩ || <2 pF

 Typical CMRR

 DC: > -80 dB
 100 kHz: > -60 dB
 3.2 MHz: > -30 dB
 100 MHz: > -26 dB

 Cable Length

 1.5 m

 

 

TOP